Press the [Tab] key to select either the wire gauge menu, cable temperature, or length (distance to Strap) for editing.

If you select the Wire Gauge menu, press the [Up] and [Down] keys to select the desired gauge.

If you choose Temperature or Length, enter the new value using the blue numeric keys. Either the Temperature or Length must be specified as unknown by entering a “#” for the value.

Note: Always enter the temperature of the cable, not the ambient temperature.

Press the [Enter] key to accept the changes and return to the previous screen. Press the [Return] key to return without saving the changes. Press the [Enter] key again to return to the hookup screen.

RFL Setup: Multi-Section

Use RFL Multi-Section when there is more than one section of cable with different wire gauges between the near-end and the far-end. Single or Separate pair hookup can be used in RFL Multi-Section.

While in the “Hookup Single Pair” or “Hookup Separate Pair” screen, press the [Right] key to select Multi Section. The wire gauge displayed near the bottom of the display will be replaced by “Multiple”. Then press the [Setup] key to go to the Setup screen.

The following is an example of the Setup screen for two sections. Note that the screen shows a “#” to indicate the distance is unknown.