Self-Test Board
A
Check Opens performance: To verify Opens, connect the red test lead to the “R” terminal, and the black test lead to the “B” terminal. Press the [Opens] key on the 965DSP. You should see the following readings for different types of cable (US and Canada).
Note: Do not connect the green test lead to the “G” terminal of the
Measurement | Type of Cable (US and Canada) | ||||
|
|
|
|
| |
|
| Capacity | Aircore | ||
|
|
|
|
| |
Red to Green | MIN | 9.35 nf | 395 ft | 350 ft | |
or | (120 m) | (107 m) | |||
|
| ||||
Black to Green |
|
|
|
| |
MAX | .011 uF | 450 ft | 400 ft | ||
| |||||
| (137 m) | (122 m) | |||
|
|
| |||
|
|
|
|
| |
Red | MIN | 9.35 nf | 595 ft | 595 ft | |
(181 m) | (181 m) | ||||
|
| ||||
to |
|
|
|
| |
Black | MAX | .011 uF | 680 ft | 680 ft | |
| (207 m) | (207 m) | |||
|
|
| |||
|
|
|
|
|
144