5.2. Digital Span (Calibration without Mass)

ESC key

If you want to return to the weighing mode during the

 

 

 

calibration mode, press the ESC key anytime. And it has

 

 

 

effect until the last displayed parameter.

ENTER key

Example: zero adjustment only, etc.

When the key is pressed, the procedure stores a current

 

 

 

parameter and proceeds to next step.

Step

1

Press and hold the ENTER key and press the

 

key to display the

 

 

 

menu in a weighing mode.

Step

2

Press the

 

key twice to select the menu CAL.

 

 

 

Press the ENTER key to enter the calibration mode.

Step

3

Press the ENTER key to enter the menu CAL.

Step

4

Select a unit using the numerical keys and press the ENTER key

 

 

to store it.

Step

5

Select a decimal point using the numerical keys and press the

 

 

ENTER key to store it.

Step

6

Select a minimum division using the numerical keys and press the

 

 

ENTER key to store it.

Step

7

Select a weighing capcity using the numerical keys and press the

 

 

ENTER key to store it.

Step

8

Press the F1 key to proceed to the digital span procedure.

Step

9

Input the zero point value using the numerical keys and press the

 

 

ENTER key to store it.

Step 10

Input the rated capacity of a loadcell using the numerical keys and

 

 

press the ENTER key to store it.

Step 11

Input the sensitivity of the loadcell in the unit of mV/V using the

 

 

numerical keys and press the ENTER key to store it.

Step 12

Press the ESC key to return the weighing mode.

Advise The digital span can be used for trimming of the actual load calibration using mass.