Page 106 AD-4402
10.4.10.4.10.4.10.4. Parameter ListParameter ListParameter ListParameter List
Category address

:

[Function] - [Function setting] - [General] - [Weighing]

Category
address
symbol Name Descriptions Range
and
choices Default
genf- 1 Display refresh rate 1: five times per second
2: ten times per second 1 to 2 2
genf- 2 Digital filter
0: Not used
1: 11 Hz
2: 8.0 Hz
3: 5.6 Hz
4: 4.0 Hz
5: 2.8 Hz
6: 2.0 Hz
7: 1.4 Hz
8: 1.0 Hz
9: 0.7 Hz
Select 3dB band for two
low pass filters. 0 to 99
4 8
genf- 3 Sampling
freqeuncy divider Use to decrease the cut-off freque ncy of
digital filter. 0 to 10 1
genf- 4 Stability detection
time 0.0 to
9.9 s1.0 s
genf- 5 Stability detection
band width
The detection condition concer ning
satblility.
0.0 : stable at anytime. 0.0 to
9.9 d2.0 d
genf- 6 Zero Range The range to zero gross display.
Center of range is zero calibration.
Unit: percentage of weighing capacity. 0 to 30 % 5 %
genf- 7 Zero tracking time 0.0 to
9.9 s0.0 s
genf- 8 Zero tracking band
width
The function automatically traces the
weighing deviation at nearly zero
point and keeps zero display of gross
display. 0.0 to
9.9 d0.0 d
genf- 9
Tare and zero
compensation at
unstable status
When unstable weghinng, whether zero
or tare command is used.
0 :Disabled
1 :Enabled
0 to 1 1
genf-10 Tare at negative
GROSS weight
When negative weghinng, whether tare
command is used.
0 :Prohibition to tare.
1 :Permission to tare.
0 to 1 1
genf-11 Preset tare Preset tare of material code
0 :Not used
1 :Use 0 to 1 1
genf-12 Preset tare =0
choice
1: If tare value of material code is zero,
the last tare value is used.
2: If tare value of material code is zero,
Tare is set to zero.
1 to 2 1
First filter