Page 100 AD-4402
9.4.9.4.9.4.9.4. Initializing ParametersInitializing ParametersInitializing ParametersInitializing Parameters
The function initializes the parameters stored in the indicator.
The parameters are stored in the flash memory and backup RAM.
Caution
There are the kinds of reset function that must re-calibrate the indicator
Note the parameters where is stored in.

Kinds of intialization mode

Kinds of intialization Description
Initializing RAM The backup RAM is reset. Zero point of gross d isplay, tare
value zeroes.
Initializing material code or
recipe code Material code and recipe code is reset.
Initializing function list It is reset parameters of function list in the flash memory.
Initializing caribration data It is reset parameters of calibration data in the flash
memory. If this function is used, calibrate the indicator.
Initializing all parameters It resets all parameters.

The location of the parameters and objects of intialization mode

Material
code Recipe
code
Zero
Tare
Material code
Accumulation data
Recipe code
Accumulation data
Function list
Calibration data
Backup RAM o o o o o o
Location Flash memory o o o o
Initializing RAM o o
Initializing material code or
recipe code o o o o
Initializing function list o
Initializing caribration data o
Intialization mode
Initializing all parameters o o o o o o o o