3.2. Connecting Loadcell Cable

Caution

Share the loadcell cable from noise-generating device and these power lines beacuse loadcell signal is sensitive.

We recommend you to use the 6 wire shielded cable to prevent loss of weighing precision.

The loadcell cable length is shorter than 5 m, you may be use a 4 wire shielded cable with terminals 1 & 2 shorted (EXC+ & SEN+ shorted) and terminals 3 & 4 shorted (EXC- & SEN- shorted).

Adaptable Compression Terminal Parts

Use the adaptable compression terminal parts to the cables

Loadcell Output Adjustment for Zero Calibration (Zero Point)

When a message "CERR2" is displayed, zero point of zero calibration is too large. When a message "CERR3" is displayed, zero point of zero calibration is too small. Use the resister more than 50 kwith low (good) temperature coefficient, when

adding a resister to adjust the loadcell output to indicator terminals.

In Case of Too Lage Output

In Case of Too Small Output