3.2. Connecting Loadcell Cable
Caution
Share the loadcell cable from
We recommend you to use the 6 wire shielded cable to prevent loss of weighing precision.
The loadcell cable length is shorter than 5 m, you may be use a 4 wire shielded cable with terminals 1 & 2 shorted (EXC+ & SEN+ shorted) and terminals 3 & 4 shorted (EXC- & SEN- shorted).
Adaptable Compression Terminal Parts
Use the adaptable compression terminal parts to the cables
Loadcell Output Adjustment for Zero Calibration (Zero Point)
When a message "CERR2" is displayed, zero point of zero calibration is too large. When a message "CERR3" is displayed, zero point of zero calibration is too small. Use the resister more than 50 kΩ with low (good) temperature coefficient, when
adding a resister to adjust the loadcell output to indicator terminals.
In Case of Too Lage Output | In Case of Too Small Output |