5.1. Actual Load Calibration (using Mass)

ESC key

If you want to return to the weighing mode during the

 

 

 

calibration mode, press the ESC key anytime. And it has

 

 

 

effect until the last displayed parameter.

ENTER key

Example: zero adjustment only, etc.

When the key is pressed, the procedure stores a current

 

 

 

parameter and proceeds to next step.

Step

1

Press and hold the ENTER key and press the

 

key to display the

 

 

 

menu in a weighing mode.

Step

2

Press the

 

key twice to select the menu CAL.

 

 

 

Press the ENTER key to enter the calibration mode.

Step

3

Press the ENTER key to enter the menu CAL.

Step

4

Select a unit using the numerical keys and press the ENTER key

 

 

to store it.

Step

5

Select a decimal point using the numerical keys and press the

 

 

ENTER key to store it.

Step

6

Select a minimum division using the numerical keys and press the

 

 

ENTER key to store it.

Step

7

Select a weighing capcity using the numerical keys and press the

 

 

ENTER key to store it.

Step

8

Perform the zero point adjustment.

 

 

Place nothing on the weighing pan and press the ENTER key to

 

 

store it after the STABLE indicator is displayed.

 

 

Whether the STABLE indicator is displayed or not, if you want to

 

 

store it, wait for ten seconds and press the ENTER key.

Step

9

Specify a total mass value to place on the weighing pan using the

 

 

numerical keys and press the ENTER key to store it.

Step 10

Place the specifyed mass on the weighing pan and press the

 

 

ENTER key to store it after the STABLE indicator is displayed.

 

 

Whether the STABLE indicator is displayed or not, if you want to

 

 

store it, wait for ten seconds and press the ENTER key.

Step 11

Press the ESC key to return the weighing mode.