AD-4402 Page 21
5.1.5.1.5.1.5.1. Actual Load Calibration (using Mass)Actual Load Calibration (using Mass)Actual Load Calibration (using Mass)Actual Load Calibration (using Mass)
ESC key If you want to return to the weighing mode during the
calibration mode, press the ESC key anytime. And it has
effect until the last displayed parameter.
Example: zero adjustment only, etc.
ENTER key When the key is pressed, the procedure stores a current
parameter and proceeds to next step.
Step 1 Press and hold the ENTER key and press the key to d isplay the
menu in a weighing mode.
Step 2 Press the key twice to select the menu CAL.
Press the ENTER key to enter the calibration mode.
Step 3 Press the ENTER key to enter the menu CAL.
Step 4 Select a unit using the numerical keys and press the ENTER key
to store it.
Step 5 Select a decimal point using the numerical keys and press the
ENTER key to store it.
Step 6 Select a minimum division using the numerical keys and press the
ENTER key to store it.
Step 7 Select a weighing capcity using the numerical keys and press the
ENTER key to store it.
Step 8 Perform the zero point adjustment.
Place nothing on the weighing pan and press the ENTER key to
store it after the STABLE indicator is displayed.
Whether the STABLE indicator is displayed or not, if you want to
store it, wait for ten seconds and press the ENTER key.
Step 9 Specify a total mass value to place on the weighing pan using the
numerical keys and press the ENTER key to store it.
Step10 Place the specifyed mass on the weighing pan and press the
ENTER key to store it after the STABLE indicator is displayed.
Whether the STABLE indicator is displayed or not, if you want to
store it, wait for ten seconds and press the ENTER key.
Step11 Press the ESC key to return the weighing mode.