5.4. Calibration Error
Error Code | Treatment and Situation |
CERR1 | Resolution (Weighing capacity / minimum division) is exceeds the limitation. |
| Increase minimum division or decrease weighing capacity. |
CERR2 | The initial load (no load output) is larger than 2mV/V. |
| Confirm the loadcell cable. |
CERR3 | Negative loadcell output value. Check wiring. |
| Confirm the loadcell cable. |
CERR4 | Mass value exceeds the weighing capacity. |
| Use a mass within the weighing capacity. (Decrease mass value) |
CERR5 | Mass value is too light for the calibration. |
| Increase mass value. |
CERR6 | The loadcell output to be equivalent to minimum division is too small. |
| Use more rough minimum division. |
CERR7 | The polarity of loadcell output is inversed. |
| Confirm the loadcell cable. |
CERR8 | The mass value of the weighing capacity exceeds 3.2 mV/V. |
| Confirm the mass and weighing capacity. |
CERR9 | Gravity acceleration is out of range. |
| Correct the value within the range of 9.770 ~ 9.835 m/s2. |
CERR10 | Zero output of loadcell unit is out of range. |
| Trim the zero output within 0.0 ~ 2.0 mV/V. |
CERR11 | The loadcell output to be equivalent to minimum division is out of range. |
| Trim the output within 0.0 ~ 3.2 mV/V. |