2.1. Precaution
Befor any use, confirm the following articles for the safty operation.
Grounding the indicator
Ground the indicator certainly. The earth terminal is the rear panel.
Separate this earth line from other ground line like a motor, inverter or a power source. Unless the indicator is grounded, it may cause to receive an electric shock, be happen operation error or catch fire
Use adaptable power cord
Confirm the AC voltage and current of the power cord. If the voltage range of cord is lower than power line voltage, it may cause of a leak or catching fire. Use compression terminals to connect the power cord to the terminal of the rear panel.
Fuse
The fuse is installed to prevent the indicator from catching fire.
The indicator is equipped many safety circuits. Therefore, the fuse is not broken in normal operation. If the fuse is broken, do not replace the fuse and contact your local or A&D dealer. This trouble may cause of an electric discharge of thunder.
Splashing water
The indicator is not
Flammability gas
Do not install the indicator in any flammable gas.
Radiation of the indicator
Space out instruments to radiate heat sufficiently.
Removing the cover
Remove the power cord terminals in the side of the power source before removing the cover to avoid receiving an electric shock.
Do not touch the internal circuit within 10 seconds after turning off the indicator because of receiving an electric shock.
Page 6 |
|