11. Specifications

General

Power supply

85 to250 VAC, 50 or 60Hz, (Stable power source)

Power consumption

Approximately 30 VA

Pysical dimensions

192 (W) x 96 (H) x 135 (D) mm

Weight

Approximately 1.8 kg

Panel cutout size

186 x 92 mm

Operation temperature

-5 to 40

Battery life of backup RAM

typ. 10 years at 25 . 5 years at 40 .

Analog to Digital Unit

Input sensitivity

Up to 0.3 V / digit

 

Zero adjustment range

0 to 2 mV /V

(0 to 20 mV)

Measuerment range

0 to 3.2 mV /V

(0 to 32 mV)

Input impedance

10 M

 

Loadcell excitation voltage

10 V DC 5%

 

Maximum loadcells

8 pieces in pallel with 350loadcell

Span temperature coefficient

8 ppm/

 

Zero temperature coefficient

0.2 V + 8 ppm/

of dead load typ.

Non-linearity

0.01 % of F. S.

 

Input noise

Below 0.3 ∝ Vp-p

 

A/D conversion

-Σconversion

 

A/D resolution

Approximately 1/1,000,000

Maximum display

16000 (to be able cancel limitation)

Sampling rate

100 times per second

Digital span function

loLoadcell offset, calibration using key operation to enter

 

the sensitivity, resolution 1/1000

Re-calibration at A/D board repacement

 

 

Omissible (resolution 1/500)

Backup method

A/D data:

EEPROM

 

Calibration:

Flash memory

 

Function:

Flash memory

 

Material code data: Backup RAM or flash memory

 

Recipe code data

Backup RAM or flash memory

Display

Main display

Fluorescent display, cobalt blue, height: 18mm, 7seg., 7 figures

Sub-display

Fluorescent display, cobalt blue, height: 5mm,

 

7seg., 54 figures and 5x7 dots, 54 figures

State indicator

Fluorescent display, cobalt blue, 8 pieces, 10 symbols, 5x7 dots

Unit indicator

Fluorescent display, cobalt blue, height: 11mm, 5x7 dots, 2 figures

Symbols

Fluorescent display, cobalt blue, height: 11mm, 5x7 dots, 2 figures

Page 132

AD-4402

Page 134
Image 134