8.2. Built-in RS-485 Interface
The
The interface can connect max. 32 units and a personal computer in a communication cable.
These unit is specified by address appended to the command.
Transmission system | EIA |
Data length | 7 bis ot 8 bits |
Start bit | 1 bit |
Parity bit | Odd, Even, not used |
Stop bits | 1 bit, 2 bits |
Baud rate | 600 bps, 12000 bps, 2400 bps, 4800 bps, 9600 bps, 19200 |
| bps, 38400 bps |
Line | 2 wires |
Connection | Max. 32 units |
Character code | ASCII code |
Terminator | CR, CR LF |
The way of Use and Note
When use terminator, connect between SDA and SDB terminals with accessory resister.
If there is not signal ground at host computer, it is not necessry to use the SG terminal.
If the shield cable is used, connect the FG teminal to shield line.
When connecting to
Use
Do not use
8.2.1.Connection
| |
A | SDA |
| |
B | SDB |
SG | SG |
| |
C.LOOP |
|
FG | Shield |
RE/DE
A
B
SG
Terminals | Interface Circuit |
Page 74 |
|