Chapter 3: Using the Logic Analyzer in Timing or State Mode

Using Symbols

To load object file symbols

Object files are created by your compiler/linker or other software development tools.

1.Generate an object file with symbolic information using your software development tools.

2.If your language tools cannot generate object file formats that are supported by the logic analyzer, create an ASCII symbol file (see page 106).

3.Select the Symbol tab and then the Object File tab.

4.Select the label name you want to load object file symbols for.

In most cases you will select the label representing the address bus of the processor you are analyzing.

5.Specify the directory to contain the symbol database file (.ns) in the field under, Create Symbol File (.ns) in This Directory. Select Browse... if you wish to find an existing directory name.

6.In the Load This Object/Symbol File For Label field, enter the object file name containing the symbols. Select Browse... to find the object file and select Load in the Browser dialog.

If your logic analyzer is NFS mounted to a network, you can select object files from other servers.

7.If your program relocates code, see “To adjust symbol values for relocated code” on page 103.

The name of the current object file is saved when a configuration file is saved. The object file will be reloaded when the configuration is loaded.

To reload object file symbols

1.Select the object file/symbol file to reload from the Object Files with Symbols Loaded For Label field.

2.Select the Reload button.

The values of the object file symbols being used in the trigger sequence or in SPA state-interval ranges will be updated automatically each time

102

Page 102
Image 102
Agilent Technologies 16760A manual To load object file symbols, To reload object file symbols, 102