Mechanical Considerations

Electrical Considerations

Chapter 2: Probing and Selecting the Sampling Mode

Probing the Device Under Test

Using the E5380A Mictor-Compatible Probe

The E5380A MICTOR-compatible probe has a MICTOR connector end. If you have a device under test with connectors designed for the Agilent E5346A high-density probe adapter, you can also use the E5380A probe.

However, the lower bandwidth of the E5380A probe limits the maximum state speed of the logic analyzer to 600 Mbits/second.

The minimum input signal amplitude required by the E5380A probe is 300 mV.

The probe matches two 17-channel probe cables to a single-ended 38 pin MICTOR connector.

Each E5380A probe requires a MICTOR connector and support shroud built into the device under test.

You can order a probing connector kit, which contains five MICTOR connectors and five support shrouds, using the Agilent part numbers:

E5346-68701 for PC board thicknesses up to 0.062 in.

E5346-68702 for PC board thicknesses from 0.062 to 0.125 in.

You can order MICTOR connectors separately using the Agilent part number:

1252-7431 (or AMP part #2-767004-2)

You can order support shrouds separately using the Agilent part numbers:

E5346-44701 for PC board thicknesses up to 0.062 in.

E5346-44704 for PC board thicknesses from 0.062 to 0.125 in.

E5346-44703 for PC board thicknesses from 0.125 to 0.70 in.

All inputs on the E5380A MICTOR-compatible probe are single-ended. The E5380A probe does not have a threshold reference voltage input. When using the E5380A probe, the logic threshold voltage is adjusted in the user interface.

39

Page 39
Image 39
Agilent Technologies 16760A manual Using the E5380A Mictor-Compatible Probe, Or AMP