Chapter 2: Probing and Selecting the Sampling Mode

Choosing the Sampling Mode

 

3. Specify when to sample. Your choices are Rising Edge, Falling Edge, or

 

Both Edges. Only Both Edges is available in the 1250 and 1500 Mb/s

 

configurations.

See Also

“To select the state speed configuration” on page 47

 

To automatically adjust sampling positions

 

When adjusting the state mode sampling position with eye finder, the

 

logic analyzer looks at signals from the device under test, figures out

 

the location of the data valid window in relation to the sampling clock,

 

and automatically sets the sampling position.

 

Because eye finder automatically runs on individual channels, it can

 

correct for the small delay effects caused by probe cables and circuit

 

board traces. This makes the logic analyzer's setup/hold window

 

smaller and lets you accurately capture data at higher clock speeds.

 

Eye finder requires:

 

• At least 500 transitions on each signal during its run. (You can use the

 

advanced eye finder settings to cause longer or shorter runs.)

 

• All devices which can drive each signal should contribute to the stimulus.

 

• All device under test operating modes relevant to the eventual logic

 

analysis measurement should contribute to the stimulus as well.

 

 

NOTE:

Eye finder measurements and normal logic analyzer measurements cannot

 

run simultaneously.

 

 

To run eye finder

1.Probe the device under test by connecting the logic analyzer channels.

2.Select the state (synchronous sampling) mode (see “To select the state mode” on page 47).

3.Format labels for those logic analyzer channels.

4.Make sure that the device under test and the logic analyzer have warmed up to their normal operating temperatures.

5.In the Format tab, select the Setup/Hold button.

49

Page 49
Image 49
Agilent Technologies 16760A manual To automatically adjust sampling positions, Automatically sets the sampling position