Chapter 3: Using the Logic Analyzer in Timing or State Mode

Setting Up Triggers and Running Measurements

See Also

“To cut/copy-and-paste sequence levels” on page 80

To cut/copy-and-paste sequence levels

You can change the order of levels in the trigger sequence by cutting- and-pasting or you can copy levels by copying-and-pasting.

1.In the Trigger tab's Trigger Sequence area, select the level that you want to cut or copy.

A yellow box appears around the level.

2.Select Cut level or Copy level from the Edit menu, or select the level button and choose Cut LEVEL or Copy LEVEL.

3.Select the level that you want to paste before or after.

4.Select Paste level before or Paste level after from the Edit menu, or select the level button and choose Paste LEVEL before or Paste LEVEL after.

See Also

“To insert/replace/delete sequence levels” on page 79

To specify a level's goto or trigger action (timing or 200 Mb/s state only)

When using multiple levels in the trigger sequence, you specify the event search order by setting the goto or trigger action in each sequence level.

1.In the Trigger tab's Trigger Sequence area, select the level whose goto or trigger action you want to specify.

A yellow box appears around the level.

2.Select the Trigger or Goto button and choose the appropriate Goto or Trigger action.

3.If you chose the Goto or Trigger and goto action, select the level button and choose the appropriate level.

Searching for events that trigger the analyzer always starts at the first level. Searching stops after one of the Trigger and fill memory actions.

One level can branch to one of several other levels depending on the

80

Page 80
Image 80