Chapter 5: Reference

The Trigger Tab

You can expand these trigger functions to see how they are constructed with the underlying advanced trigger functions.

You can break down these trigger functions to directly edit the underlying advanced trigger functions. Sometimes you need to break down a trigger function in order to add other event types such as counters, timers, flags, and arming from other instruments.

Find pattern n times

Becomes true when the specified pattern occurs in the specified number of samples (eventually).

Store range until pattern occurs

Becomes true when the specified pattern occurs in the specified number of samples (eventually) and only stores samples in the specified range until then.

Store pattern2 until pattern1 occurs

Becomes true when the first specified pattern occurs in the specified number of samples (eventually) and only stores samples with the second specified pattern until then.

While storing pattern2, find pattern1

This trigger function has been replaced by the "Store range until pattern occurs" and "Store pattern2 until pattern1 occurs" trigger functions.

Store nothing until pattern occurs

Becomes true when the specified pattern occurs one time and doesn't store any samples until then.

Run until user stop

Sets up to never trigger. You must select the stop button to view the captured data.

Find pattern2 occurring immediately after pattern1

Becomes true when the second specified pattern occurs in the sample immediately after a sample in which the first specified pattern occurs.

Find pattern1 eventually followed by pattern2

181

Page 181
Image 181
Agilent Technologies 16760A manual Find pattern1 eventually followed by pattern2, 181