57
Chapter 2: Probing and Selecting the Sampling Mode
Formatting Labels for Logic Analyzer Probes
Formatting Labels for Logic Analyzer Probes
The Format tab is mainly for assigning bus and signal names (from the
device under test), to logic analyzer channels. These names are called
labels. Labels are used when setting up triggers and displaying
captured data.
The Format tab also lets you do things like assign pods to the logic
analyzer, specify the logic analyzer threshold voltage, change the label
polarity, reorder bits in a label, and turn labels off or on.
The Format tab has activity indicators that show signal levels.
“To assign pods to the analyzer” on page57
“To set pod threshold voltages” on page 58
“To set clock threshold voltages” on page 59
“To assign probe channels to labels” on page 60
“To import label names and assignments from a netlist” on page 62
“To import label definitions from an ASCII file” on page 63
“To export label definitions to an ASCII file” on page 64
“To change the label polarity” on page 64
“To reorder bits in a label” on page 65
“To turn labels off or on” on page 66

To assign pods to the analyzer

The logic analyzer pods can be assigned to the logic analyzer, or, they
can be left unassigned.
1. In the Format tab, select the Pod Assignment button.
2. In the Pod Assignment dialog, drag a pod to the logic analyzer.
3. Select the Close button.