222
Chapter 5: Reference
Error Messages
The clock's threshold level is set by the po d threshold. For the logic
analyzer's J clock, check the pod threshold of pod 1 of the master card.
Timer value checked as an event, but no start action specified
This warning occurs because you have used a timer in your trigger
sequence, but have not started it with either Start from reset or
Resume in any action. You do not need to start the timer in the same
sequence level. The timer will still function if not started, but will not
change value.
Trigger function initialization failure
The "trigger function initialization failure" messages mean that you
tried to insert a trigger function which required a change to the trigger
function default state.
Possible Causes
Tried to insert "Wait for arm in" trigger function
A "Wait for arm in" trigger level causes the logic analyzer to wait for a
signal from another module or Port In. These signals are passed through
the Intermodule Bus. To prevent the logic analyzer from hanging, it must
be added to the Group Run Arming Tree. To do this, open the Intermodule
window by selecting the Intermodule icon in the System window. In the
Intermodule window, select the analyzer icon and select any option except
for Independent.
Tried to insert "Wait for other machine to trigger" function
A "Wait for the other machine to trigger" trigger level causes the logic
analyzer to wait for a signal from the other logic analyzer machine in the
module. If this machine is not on, the current logic analyzer will ha ng. To
turn the other machine on, select Pod Assignment under Format. Set the
type of the other machine to State or Timing.
•In Format, no labels have bits assigned to them.