Chapter 5: Reference

The Trigger Tab

When the logic analyzer is armed by another instrument (as specified in the Intermodule window), this trigger function becomes true when the arm signal is received.

Wait for flag

Becomes true when the specified flag has the specified value. This trigger function tests for a flag event.

Set/clear/pulse flag

Becomes true on any sample and sets, clears, pulse sets, or pulse clears the specified flag. This trigger function inserts a flag action.

OR Trigger

When the logic analyzer is armed by another instrument (as specified in the Intermodule window), this trigger function becomes true when a pattern occurs a specified number of times OR when the arm signal is received.

“Advanced Trigger Functions” on page 183

See Also

“To specify a label pattern event (Timing only)” on page 71

 

“To break down a trigger function (timing or 200 Mb/s state only)” on

 

page 72

 

“To cross-trigger with another instrument” on page 112

Advanced Trigger Functions

The advanced trigger functions let you create a custom trigger sequence level using events, comparison functions, and up to 4 branches.

The advanced trigger functions are available in the timing sampling mode and in the state sampling mode's 200 Mb/s / 32M State configuration. In the timing sampling mode, you can specify edges on labels and event durations.

The types of events include labels, timers, flags, and counters.

• Advanced - If/then

This trigger function has only one branch. If the events in the "If" event list