59
Chapter 2: Probing and Selecting the Sampling Mode
Formatting Labels for Logic Analyzer Probes
appropriate threshold voltage reference level.
Select the Differential option. This option appears when the E5379A
differential probe is used. It should be selected when differential
signals are probed. The difference voltage (Vin+ - Vin-) must be greater
than or equal to 200 mV p-p.
Select the User Defined option and enter the desired threshold voltage
value. The threshold level is selectable from -6.0 volts to +6. 0 volts.
3. If you don't want the change to apply to all pods and clock inpu t
thresholds, deselect the checked box next to Apply threshold setting to
all pods.
4. Select the Close button.
NOTE: The logic analyzer requires a minimum vol tage swing of 250 mV for the
E5378A single-ended probe or 300 mV for the E5380A MICTOR-compatib le
probe to recognize changes in logic levels. If you are using the E5379A
differential probe, 200 mV differential is required.
NOTE: The specified pod threshold voltage is also applied to the pod's clock
threshold if Apply settings to all pods is selected. However, the pod's clock
threshold can also be changed independently.
See Also “To set clock threshold voltages” on page 59
“Using the E5378A Single-Ended Probe” on page 35
“Using the E5379A Differential Probe” on page37
“Using the E5380A Mictor-Compatible Probe” on page 39
“Using the E5382A Single-ended Flying Lead Probe Set” on p age 40
To set clock threshold voltages

The threshold voltage is the voltage level that a signal must cross

before the logic analyzer recognizes a change in logic levels.

1. In the Format tab, selec t the Clk Thresh... button located just below Data
On Clocks.