Chapter 3: Using the Logic Analyzer in Timing or State Mode

Setting Up Triggers and Running Measurements

page 72

“Saving/Recalling Trigger Setups” on page 90

Using Other Trigger Features

Other subtabs in the Trigger tab let you do things like: specifying whether a state or time count is stored with samples, or setting up the default storing options.

“To count time, states, or turn counting off” on page 75

“To specify default storing” on page 76

“To specify whether default storing is initially on or off” on page 78

To count time, states, or turn counting off

You can specify whether a time or state count is stored with samples.

1.In the Trigger tab's Settings subtab, select the Count option button and choose either Off, Time, or States.

The States option is only available in the state sampling mode's 200 Mb/s / 32M State configuration.

2.If you chose States:

a.Select the Define button.

b.In the State count qualify dialog, select the Count if or Count if NOT option.

c.Specify events that identify the states to be counted or not counted.

d.If you would like to specify the evaluation order of the event list, select Group events. Then, in the Group Events dialog, either select the Add parens button to group events or select the Remove parens button to ungroup events. When you're done grouping events, select the OK button.

In the state mode's 200 Mb/s / 32M State configuration, memory depth is reduced by half if all pods are used and time or state counts are being stored. In all other state mode configurations, time counts require one

75

Page 75
Image 75
Agilent Technologies 16760A manual Using Other Trigger Features, To count time, states, or turn counting off