Chapter 5: Reference

Error Messages

Maximum of 32 Channels Per Label

The logic analyzer can only assign up to 32 channels for each label. If you need more than 32 channels, assign them to two labels and use the labels in conjunction.

Must assign another pod pair to specify actions for flags

In state sampling mode, when there is only one pod pair assigned to an analyzer, flags are not available. You must assign another pod pair to the analyzer in order for flags to be available.

In the timing sampling mode, flags are are always available.

“To assign pods to the analyzer” on page 57

Must assign Pod 1 on the master card to specify actions for flags

When using a 16760A analyzer in 200Mb/s state mode, Pod 1 on the master card must be assigned in order to add actions for the flags in a branch action list. To assign the pod, go to the format tab and select the "Pod Assignment" button. Drag Pod 1 of the master card from the unassigned pods list to the analyzer assigned list.

No more Edge/Glitch resources available for this pod pair

This error occurs when you have used more than 2 edges or glitches per pod pair in the trigger specification.

219