Chapter 5: Reference

The Sampling Tab

 

channel will select it in each of those labels.

See Also

“Understanding State Mode Sampling Positions” on page 256

 

“To automatically adjust sampling positions” on page 49

 

Eye Finder Advanced Settings Dialog.

Short

Eye finder looks at 100,000 clock cycles on each channel

 

to determine the suggested sampling positions. This setting

 

requires frequent transitions on all channels.

Medium

Eye finder looks at 500,000 clock cycles on each channel

 

to determine the suggested sampling positions. Use this for

 

channels that transition at a normal rate.

Long

Eye finder looks at 2.5 million clock cycles on each

 

channel to determine the suggested sampling positions.

 

Use this setting if some channels have sporadic transitions.

Custom

Lets you specify the number of clock cycles to sample on

 

each channel at each sampling position (each point in

 

time).

Accumulate

 

Results

When selected, eye finder data is accumulated from run-

 

to-run. When unselected, eye finder data is erased before

 

each run.

Some things to consider when selecting among the eye finder advanced settings are:

Upper address bits that don't transition as frequently as lower address bits.

170

Page 170
Image 170
Agilent Technologies 16760A manual Eye Finder Advanced Settings Dialog, 170