Chapter 5: Reference

The Trigger Tab

captured data. For maximum data capture the trigger position should be set to end.

A warning message concerning no trigger action will be displayed in the Run Status window. This warning message can be ignored since the intent of the trigger function is not to trigger. The run status window can be prevented from coming up with this warning by setting the Popup On Severity level to errors only.

16760 Half Channel State Trigger Functions

When the state sampling mode's 1250 Mb/s / 128M Half Chan or 1500 Mb/s / 128M Half Chan configuration is selected, the following 16760 Half Channel State trigger functions are found in the Trigger Functions tab.

You can edit these trigger functions to specify particular pattern events or boolean combinations of events.

You can use the following types of events in these trigger functions: patterns, ranges, flags, and Wait for arm in (see “To cross-trigger with another instrument” on page 112).

These trigger functions provide trigger actions. You cannot modify these actions or insert any other actions.

Only one of these trigger functions can be used at a time; in other words, they cannot be sequenced to provide more sophisticated triggering.

Find pattern

Trigger when the specified pattern occurs.

Find 2 patterns in immediate sequence

Trigger when the second specified pattern occurs in the next sample after a sample in which the first specified pattern occurs.

Find 2 patterns in eventual sequence

Trigger when the second specified pattern occurs, eventually, after the first specified pattern has already occurred.

Run until user stop