Chapter 4: Using the Logic Analyzer in Eye Scan Mode

Displaying Captured Eye Scan Data

3. Select the Linear button to equally divide the color scale so that each

 

color represents an equal increment of transitions.

 

Select the Log, Log2, or Log10 scale buttons to separate the display of

 

transitions detected using a logarithmic scale.

 

4. Use the Parts per Million (ppm) slider bars above and below the color

 

bars to bring out detail in the eye diagram. The bottom slider bar controls

 

the black level. Set the slider bar or enter a number in the text box

 

associated with the black color bar to hide noise. The upper limit for black

 

is zero by default.

 

 

NOTE:

Measurement points that are excluded from the display using the value

 

entered for the black color bar will not be reported by the measurement tools.

 

To cause the measurement tools to report all measurement points, enter 0

 

(zero) in the box for the black color bar in the Color Scale window. See “To

 

make measurements on the eye scan data” on page 142.

 

Similarly, you can use the top slider bar or enter a number in the text

 

141