Chapter 3: Using the Logic Analyzer in Timing or State Mode

Setting Up Triggers and Running Measurements

 

• “To clear the trigger sequence” on page 83

See Also

“Sequence Levels” on page 242 in “Understanding Logic Analyzer

 

Triggering” on page 240

To insert/replace/delete sequence levels

To insert sequence levels

1. In the Trigger tab's Trigger Sequence area, select the level that you want to insert before or after.

A yellow box appears around the level.

2. In the Trigger Functions subtab, select the trigger function you want to insert.

A picture describing the trigger function is shown.

3. Select the Insert before or Insert after button, or select the level button and choose Insert LEVEL before or Insert LEVEL after.

To replace sequence levels

1.In the Trigger tab's Trigger Sequence area, select the level that you want to replace.

A yellow box appears around the level.

2.In the Trigger Functions subtab, select the trigger function you want to insert.

A picture describing the trigger function is shown.

3.Select the Replace button, or select the level button and choose Replace

LEVEL.

To delete sequence levels

1.In the Trigger tab's Trigger Sequence area, select the level that you want to delete.

A yellow box appears around the level.

2.Select the Delete button, or select the level button and choose Delete

LEVEL.