Chapter 5: Reference

The Sampling Tab

 

as the sampling clock.

 

Generally, the state mode sampling clock is taken from the

 

signals that clock valid data in the device under test.

See Also

“Selecting the State Mode (Synchronous Sampling)” on page 46

 

“In Either Timing Mode or State Mode” on page 53

Sampling Positions Dialog

The Sampling Positions dialog lets you position the logic analyzer's setup/hold window (or sampling position) so that data on high-speed buses is captured accurately, in other words, so that data is sampled when it is valid.

When the device under test's data valid window is less than 2.5 ns (roughly, for clock speeds >= 200 MHz), it's easiest to use eye finder to locate the stable and transitioning regions of signals and to automatically adjust sampling positions.

When the device under test's data valid window is greater than 2.5 ns (roughly, for clock speeds < 200 MHz), it's easiest to adjust the sampling position manually, without using the logic analyzer to locate the stable and transitioning regions of signals.

 

• “Sampling Positions Tab” on page 160

 

• “Eye Finder Run Messages” on page 162

 

• “Eye Finder Info Messages” on page 165

 

• “Eye Finder Load/Save Messages” on page 167

 

• “Eye Finder Setup Tab” on page 169

 

• “Eye Finder Advanced Settings Dialog” on page 170

 

• “File Info Tab” on page 171

See Also

“Understanding State Mode Sampling Positions” on page 256

 

“Selecting the State Mode (Synchronous Sampling)” on page 46