Chapter 3: Using the Logic Analyzer in Timing or State Mode

Setting Up Triggers and Running Measurements

(label events), and the action is to trigger the logic analyzer. However, events can also test timer, counter, and/or flag values that are set up in the logic analyzer, and actions can include setting up timers, counters, and flags as well as specifying special store actions.

“To specify a duration or occurrence count for events” on page 84

“To insert a store action (state mode)” on page 85

“To insert timer actions/events” on page 85

“To insert counter actions/events” on page 86

“To insert flag actions/events” on page 87

“To insert a "reset occurrence counter" action” on page 89

“To group events” on page 89

“To use named events” on page 89

To specify a duration or occurrence count for events

When working with advanced trigger functions or when you break down other trigger functions, you can specify that an event be present for a certain amount of time, or occur in a certain number of samples, before the associated action is taken.

To specify a time duration for events (timing only)

1.In the Trigger tab's Trigger Sequence area, if the present for > button is not present, select the occurs button and choose present for >.

2.Enter a time duration value.

The event must be present for the specified period of time before the action is taken.

To specify a < duration, break down the Find pattern present for < duration trigger function. (It actually uses occurrence counts and four sequence levels.)

To specify an occurrence count for events (timing, 200, 400 Mb/s only)

1.In the Trigger tab's Trigger Sequence area, if the occurs button is not present, select the present for > button and choose occurs.