Chapter 4: Using the Logic Analyzer in Eye Scan Mode

Setting Up and Running Eye Scan Measurements

 

6.

Select Sampling Positions... to open the Sampling Positions dialog. Select

 

 

the label you created and choose Expand to display all the signals (if it is

 

 

not already expanded).

 

7.

Start your target system and select Run Eye Finder.

 

8.

When Eye Finder completes its run, move the qualifier sampling position

 

 

to the right of the qualifier transition region at or to the right of the clock

 

 

position (t = 0). The sampling position is shown by the blue line in the

 

 

display.

 

 

NOTE:

The sampling positions for data are not used by eye scan. They should be set

 

for correct state mode acquisitions.

 

 

 