Example 3

Programming

Programming Examples

Example 3 applies to measurements where the 85330A controls the sources, using Fast Source Control mode.

1001

!

 

1002

! RE-SAVE “EX3”

 

1003

!

 

1004

Example_3:!

 

1005

!

 

1006

! This example shows how to make multi-frequency, multi-channel

1007

! measurements using the HP 85330A and the FAST DATA modes of the HP 8530A

1008

! Microwave Receiver. In this mode the sources are setup by the computer

1009

! rather than under control of the HP 8530A microwave receiver.

1010

!

 

1011

! It uses the HP BASIC/WS TRANSFER command to read data from the receiver

1012

! The HP BASIC command ENTER may also be used.

1013

!

 

1014

DIM Err_str$[128]

 

1015

INTEGER Data_f1(1:32000,0:2) BUFFER

! DATA FROM RECEIVER IN FORMAT FORM1.

1016

!

 

1017

Build_table:!

 

1018

!

 

1019

ALLOCATE REAL Exp_tbl(0:255)

 

1020

!

 

1021

! Build the FORM1 to FORM3 coversion table. During the FAST DATA

1022

! aquisition from the HP 8530A data translation is need to conver the

1023

! 6 bytes read from the receiver to a data point consiting of a

1024

! real and imaginary pair.

 

1025

!

 

1026

Exp_tbl(0)=2^(-15)

 

1027

FOR N=0 TO 126

 

1028

Exp_tbl(N+1)=Exp_tbl(N)+Exp_tbl(N)

 

1029

NEXT N

 

1030

Exp_tbl(128)=2^(-143)

 

1031

FOR N=128 TO 254

 

1032

Exp_tbl(N+1)=Exp_tbl(N)+Exp_tbl(N)

 

1033

NEXT N

 

1034

!

 

1035

Set_vars:!

 

1036

!

 

1037

! Set variables

 

1038

!

 

1039

Scu_addr=0

! SCU address

1040

Chan_start=1

! First Channel

1041

Chan_stop=4

! Last Channel

1042

Chan_pts=(Chan_stop-Chan_start)+1

! Number of channels

1043

Switch_start=Scu_addr*100+Chan_start

! Starting SWITCH ADDRESS

1044

Switch_stop=Scu_addr*100+Chan_stop

! Ending SWITCH ADDRESS

85330A Multiple Channel Controller