Default
Syntax
Example
NOTE
Default
Syntax
Example
NOTE
Remote Programming Command Reference
Standard Command Reference
ROUTe:CONTrol
Controls the TTL state of the rear panel AUX 1 connector.
Default is a TTL LOW.
OUT:CONT <data>;
ROUT:CONT 0;
Sets AUX 1 to TTL LOW
ROUT:CONT 1;
Sets AUX 1 to TTL HIGH
The state of AUX 1 must be set before entering
ROUTe:DELay
Adds a specific amount of delay between a switch address entry being asserted and the issuance of a measurement trigger. Delay is entered in
The default is 160
ROUT:DEL
ROUT:DEL 200;
Set the trigger delay (settling time) to 200 micro-seconds.
Most solid state switches are capable of changing states within a few
85330A Multiple Channel Controller |