Default

Syntax

Example

NOTE

Default

Syntax

Example

NOTE

Remote Programming Command Reference

Standard Command Reference

ROUTe:CONTrol

Controls the TTL state of the rear panel AUX 1 connector.

Default is a TTL LOW.

OUT:CONT <data>;

ROUT:CONT 0;

Sets AUX 1 to TTL LOW

ROUT:CONT 1;

Sets AUX 1 to TTL HIGH

The state of AUX 1 must be set before entering run-time mode.

ROUTe:DELay

Adds a specific amount of delay between a switch address entry being asserted and the issuance of a measurement trigger. Delay is entered in micro-seconds.

The default is 160 micro-seconds.

ROUT:DEL <micro-seconds>;

ROUT:DEL 200;

Set the trigger delay (settling time) to 200 micro-seconds.

Most solid state switches are capable of changing states within a few micro-seconds.

7-6

85330A Multiple Channel Controller

Page 116
Image 116
Agilent Technologies 85330A service manual ROUTeCONTrol, ROUTeDELay