Default
Syntax
Example
NOTE
Default
Syntax
Example
NOTE
Remote Programming Command Reference
Standard Command Reference
RUNTime:SOURce:SOURCE1:TRIGger
Selects IMM, TTL, REMote1, or REMote2 triggering for source 1 (the RF source).
The default trigger mode is TTL.
RUNTime:SOURce:SOURCE1:TRIGger IMM TTL
REMote1 REMote2;
RUNT:SOUR:SOURCE1:TRIG IMM;
In the fast source stepping mode (also known as Turbo), the trigger is set to TTL. When using more than one multiple channel controller, the CONTroller controller would set the source trigger parameter to remote (REMote1 or REMote2), while the REMote1 or REMote2 controller sets its parameter to TTL.
RUNTime:SOURce:SOURCE2:TRIGger
Selects IMM or TTL triggering for source 2 (the LO source).
The default trigger mode is TTL.
RUNTime:SOURce:SOURCE2:TRIGger IMM TTL
REMote1 REMote2;
RUNT:SOUR:SOURCE2:TRIG IMM;
In the fast source stepping mode (also known as Turbo), the trigger is set to TTL. When using more than one multiple channel controllers, the CONTroller controller would set the source trigger parameter to remote (REMote1 or REMote2), while the REMote1 or REMote2 controller sets its parameter to TTL.
85330A Multiple Channel Controller |