AUX 1 and AUX 2 Output Voltage Test
When Finished with All Tests
Performance Verification
AUX 1 and AUX 2
The following test will determine if the rear panel ports successfully output the correct DC switch drive levels.
1. On the oscilloscope press:
[RECALL] [1]
[TRIG] {AUTO}
[CLEAR DISPLAY]
2. Press the following on the multiple channel controller:
[Clear Instr] [Select Instr] {85330A}
3.Connect a BNC cable between the AUX 1 connector and channel 1 of the oscilloscope.
a. Press:
{CLOSE} {AUX1} {0} or [f1]
The oscilloscope should show a trace of about 0 Vdc.
b. Press:
{1} or [f2]
The oscilloscope should show a trace of about 4 Vdc.
4.Connect a BNC cable between the AUX 2 connector and channel 1 of the oscilloscope.
a. Press:
[Clear Instr] {CLOSE} {AUX2} {0} or [f1]
The oscilloscope should show a trace of about 0 Vdc.
b. Press:
{1} or [f2]
The oscilloscope should show a trace of about 4 Vdc.
All of the performance verification tests are now complete. Reconnect the 85330A system in the original measurement system configuration. Refer to Chapter 1, “Installation” for more information.
85330A Multiple Channel Controller |