Agilent Technologies 85330A service manual Digital Oscilloscope, Procedure

Models: 85330A

1 166
Download 166 pages 5.42 Kb
Page 34
Image 34

 

Performance Verification

 

Performance Verification Test

Digital Oscilloscope

Set up the Agilent 54503A digital oscilloscope (or any oscilloscope) as

 

described in the following procedure, using the same settings.

 

 

 

 

 

 

NOTE

The oscilloscope setup described in the following procedure will not display

 

a trace on the oscilloscope screen until a pulse is initiated. The oscilloscope

 

will then hold the trace until the [CLEAR DISPLAY] is pressed.

 

Procedure

 

 

Turn ON the oscilloscope and press the following keys on the 54503A:

 

[CHAN]: {1}

 

{ON}

 

{2v/Div}

 

{Offset: 0v}

 

{Coupling: DC}

 

{Input: 1 M }

 

{Probe Ratio: 1:1}

 

[CHAN]: {2}

 

{ON}}

 

{2v/Div}

 

{Offset: 5v}

 

{Coupling: DC}

 

{Input: 1 M }

 

{Probe Ratio: 1:1}

 

[TRIG]:

 

{Trig’d}

 

{Edge}

 

{Source: 1}

 

{Adjust: 1v}

 

{

 

 

 

} (- Edge)

 

 

 

 

[TIMEBASE]:

 

{2 us/Div}

 

{Delay: 0}

 

{Reference: Cntr}

 

[SAVE] [1]

2-4

85330A Multiple Channel Controller

Page 34
Image 34
Agilent Technologies 85330A service manual Digital Oscilloscope, Procedure