Performance Verification
Verifying the Multiple Channel Controller
Verifying the Multiple Channel Controller
Turn-On
This portion of the verification will check the 85330 multiple channel switch controller to ensure that it is operating correctly. It will also check the Switch Control Unit of the Agilent 85331/2 to make sure that it is switching correctly. Document the test results by plotting or printing the test instrument screens as each test is performed.
Turn the multiple channel controller ON; the following screen should appear:
Select an instrument._
| SYSTEM 85330A IBASIC (see note below) |
|
|
NOTE | IBASIC is only present if the IBASIC option is installed. |
| If you see the above display, the unit has passed its turn on tests and is |
| |
| working properly. If you do not see this display, load the 85330A |
| downloadable driver files as explained in the “Loading the Driver from a |
| Personal Computer” on page |
Voltage Check
Check the voltage on the switch control unit’s OUTPUT connector. This is the connector that is used to
Table 2-2 Switch Control Unit Voltages
Pin | Voltage |
|
|
E, L, V | +12 ±1.8 Vdc |
|
|
a | −12 ±1.8 Vdc |
|
|
F, M, W, b | ground |
|
|
85330A Multiple Channel Controller |