Counter Pulse Delay Test
Performance Verification
Verifying the Multiple Channel Controller
The following test determines if the
1.Connect BNC cables between the SRC1 TRIG, RCVR READY connectors and the oscilloscope as shown in Figure
2.On the oscilloscope press:
[RECALL] [1]
[TRIG] {Source 2} {Adjust: 1.5v} { } (+ Edge}
[TIMEBASE] {1 us/Div} {Delay: 4 us} {Ref: Centr}
[SAVE] [2]
3. Press the following on the multiple channel controller:
[Clear Instr] [Select Instr] {85330A} {RESET} [More] {TEST} {COUNTER}
4.Press the following on the multiple channel controller:
[4] [Return]
5.The negative edge of the displayed pulse should be 4 ±1 microseconds.
To use the scope measurement feature, press: [∆ t ∆ v] {∆ t Marker: ON}. Rotate the front panel knob until the vertical marker line intersects the negative edge of the pulse. Note that the Start Marker reading will be the pulse delay. See Figure
6.On the oscilloscope press:
[TIME BASE] {1 us/Div} {Delay: 1000 us} [CLEAR DISPLAY]
7.Press the following on the multiple channel controller:
[1] [0] [0] [0] [Return]
8.The negative edge of the displayed pulse should be 1,000 ± 1 microseconds.
To use the scope measurement feature, press: [∆ t ∆ v] {∆ t Marker: ON}. Rotate the front panel knob until the vertical marker line intersects the negative edge of the pulse. Note that the Start Marker reading will be the pulse delay. See Figure
9.On the oscilloscope press:
[TIME BASE] {2 us/Div} {Delay: 15 ms} [CLEAR DISPLAY]
10.Press the following on the multiple channel controller:
[1] [5] [0] [0] [0] [Return]
11.The negative edge of the displayed pulse should be 15 ± 0.001 milliseconds.
To use the scope measurement feature, press: [∆ t ∆ v] {∆ t Marker: ON}. Rotate the front panel knob until the vertical marker line intersects the
85330A Multiple Channel Controller |