Agilent Technologies 85330A Network Analyzer Cables Calibration Save Setup, Measurement Display

Models: 85330A

1 166
Download 166 pages 5.42 Kb
Page 35
Image 35

Network Analyzer

Cables

Calibration

Save Setup

Performance Verification

Performance Verification Test

Turn ON the vector network analyzer. Adjust the controls for the following setup:

Measurement Display

Display:

4 Channel

 

Channel 1:

S11

SWR

Channel 2:

S21

Log Mag

Channel 3:

S12

Log Mag

Channel 4:

S22

SWR

Domain:

 

Frequency

Start Frequency:

45 MHz

Stop Frequency:

40 GHz

Number of Points:

201

RF Source Power:

Maximum

Measurement:

 

Continuous

Averaging:

 

As required

Set up the rest of the measurement display for your convenience and ease of viewing.

Connect one test port cable to Port 2 of the test set. The RF switch will be connected directly to the test set Port 1. An example of this setup is shown in Figure 2-1, “An Example Performance Verification Setup,” on page 2-3.

Perform a full, two port calibration at the test set Port 1, and at the end of the test port cable connected to Port 2.

Save this instrument state setup in Save/Recall register 1. It will be used later in the tests. Press:

[SAVE] {1}

85330A Multiple Channel Controller

2-5

Page 35
Image 35
Agilent Technologies 85330A service manual Network Analyzer Cables Calibration Save Setup, Measurement Display