Network Analyzer
Cables
Calibration
Save Setup
Performance Verification
Performance Verification Test
Turn ON the vector network analyzer. Adjust the controls for the following setup:
Measurement Display
Display: | 4 Channel  | 
  | 
Channel 1:  | S11  | SWR  | 
Channel 2:  | S21  | Log Mag  | 
Channel 3:  | S12  | Log Mag  | 
Channel 4:  | S22  | SWR  | 
Domain:  | 
  | Frequency  | 
Start Frequency:  | 45 MHz  | |
Stop Frequency:  | 40 GHz  | |
Number of Points:  | 201  | |
RF Source Power:  | Maximum  | |
Measurement:  | 
  | Continuous  | 
Averaging:  | 
  | As required  | 
Set up the rest of the measurement display for your convenience and ease of viewing.
Connect one test port cable to Port 2 of the test set. The RF switch will be connected directly to the test set Port 1. An example of this setup is shown in Figure 
Perform a full, two port calibration at the test set Port 1, and at the end of the test port cable connected to Port 2.
Save this instrument state setup in Save/Recall register 1. It will be used later in the tests. Press:
[SAVE] {1}
85330A Multiple Channel Controller  |