Network Analyzer

Cables

Calibration

Save Setup

Performance Verification

Performance Verification Test

Turn ON the vector network analyzer. Adjust the controls for the following setup:

Measurement Display

Display:

4 Channel

 

Channel 1:

S11

SWR

Channel 2:

S21

Log Mag

Channel 3:

S12

Log Mag

Channel 4:

S22

SWR

Domain:

 

Frequency

Start Frequency:

45 MHz

Stop Frequency:

40 GHz

Number of Points:

201

RF Source Power:

Maximum

Measurement:

 

Continuous

Averaging:

 

As required

Set up the rest of the measurement display for your convenience and ease of viewing.

Connect one test port cable to Port 2 of the test set. The RF switch will be connected directly to the test set Port 1. An example of this setup is shown in Figure 2-1, “An Example Performance Verification Setup,” on page 2-3.

Perform a full, two port calibration at the test set Port 1, and at the end of the test port cable connected to Port 2.

Save this instrument state setup in Save/Recall register 1. It will be used later in the tests. Press:

[SAVE] {1}

85330A Multiple Channel Controller

2-5