Remote Programming Command Reference

Common Commands

*OPC?

Syntax

Example

The operation complete command. This can be used to measure timing of a run-time sequence. The command *OPC? returns a value of 1 when evoked. Prior commands are executed before the completion of *OPC?.

*OPC?

10 OUTPUT 70918;”RUNT:EVEN:COUN 1;”

20 OUTPUT 70918;”TRIG:SOUR IMM;”

30 TIME_START=TIMEDATE

40 OUTPUT 70918;”INIT:IMM;*OPC?”

50 ENTER 70918;OPC_NUM

60 TIME_STOP=TIMEDATE

70 PRINT “MEASUREMENT TIME (SEC)=”;TIME_STOP-TIME_START

 

*RST

 

A soft reset of the E1330A/B and 85330-60002 card. It is executed at

 

power-up. This command does not clear the error stack.

Default

All card address are set to channel 1.

Syntax

*RST

Example

OUTPUT 70918;”*RST”

 

 

NOTE

A *RST will produce a trigger signal on the SRC 1 TRIG, SRC2 TRIG, and

 

RCVR TRIG lines.

 

 

7-4

85330A Multiple Channel Controller

Page 114
Image 114
Agilent Technologies 85330A service manual Opc?, Default