Installation

Selecting Positive or Negative-Edge Event Triggers

Figure 1-4 Positive/Negative-Edge EVENT TRIG jumpers

1.Remove the two screws that hold the E1330B board in place.

2.Remove the four screws that hold the 85330-60002 board in place.

3.Pull out the two boards at the same time. You can do this by pulling each one out a little at a time, until they are both removed. Place the boards on an anti-static surface.

Standard positive-edge operation is set at the factory by jumpering E9 and E10 together.

4.To select negative-edge operation, move the jumper across E9 and E8.

5.Reinstall the boards and screws.

85330A Multiple Channel Controller

1-9

Page 27
Image 27
Agilent Technologies 85330A service manual Positive/Negative-Edge Event Trig jumpers