Programming

 

 

Programming Examples

1089

OUTPUT @Rec;”TRGSEXT;HOLD;” ! SET TO EXTERNAL TRIGGER FOR CONTROLLED

1090

!

 

 

1091

!

 

 

1092

Set_85330a:!

 

 

1093

!

 

 

1094

OUTPUT @Hp85330a;”*RST;”

 

! SOFT RESET

1095

!

 

 

1097

OUTPUT @Hp85330a;”RUNT:EVEN:TRIG TTL;”

 

! Set the triggering.

1098

OUTPUT @Hp85330a;”RUNT:EVEN:COUN “;Angle_pts;”;”

! Angle increments

1099

!

 

 

1100

OUTPUT @Hp85330a;”RUNT:TIME:EVEN 0,15000000;”

! timeout 1st point: 15 sec.

1101

OUTPUT @Hp85330a;”RUNT:TIME:EVEN 1,5000000;”

! timeout 2nd - last: 5 sec.

1102

OUTPUT @Hp85330a;”RUNT:TIME:REC 1000000;”

 

! timeout receiver: 1 sec.

1103

!

 

 

1104

OUTPUT @Hp85330a;”RUNT:SWIT:DEL 50;”

 

! Switch settling is 50 uS

1105

!

 

! when using multi-parameter

1106

!

 

! display.

1107

OUTPUT @Hp85330a;”RUNT:SWIT:SCAN 1,(@”;Switch_start;”:”;Switch_stop;”);”

1108

!

 

 

1109

OUTPUT @Hp85330a;”RUNT:SOUR:COUN “;Freq_pts;”;”

! Frequency points.

1110

OUTPUT @Hp85330a;”RUNT:SOUR:SOURCE1:TRIG IMM;”

! No External triggering.

1111

OUTPUT @Hp85330a;”RUNT:SOUR:SOURCE2:TRIG IMM;”

! No External triggering.

1112

!

 

 

1113

!

 

 

1114

Set_positioner:!

 

 

1115

!

 

 

1116

! Set positioner to start angle.

 

 

1117

! Set velocity, acceleration.

 

 

1118

! Set start, stop and increment angles.

 

 

1119

! Set positioner to issue TTL triggers at increment angles.

1120

!

 

 

1121

!

 

 

1122

Start_meas:!

 

 

1123

!

 

 

1124

OUTPUT @Rec;”CLES;SING;”

! Start receiver single sweep.

1125

OUTPUT @Hp85330a;”RUNT:INIT:IMM;”

! Start measurement process

1126

REM Set the positioner to take an angle scan. ! Start the positioner.

1127

!

 

 

1128

FOR Passes=1 TO Angle_pts

 

 

1129

REPEAT

 

 

1130

! Wait for SING sweep to complete

 

 

1131

UNTIL BIT(SPOLL(@Rec),4)

 

 

1132

FOR N=1 TO Chan_pts

 

 

1133

OUTPUT @Rec;”PARA”&VAL$(N)&”;”

 

 

1134

OUTPUT @Rec;”FORM3;OUTPDATA;”

 

 

85330A Multiple Channel Controller