Glossary
Standard System
A “standard system” uses one 85330A with one or two switch control units, each with a
•If you have two SCUs, both can use the default SCU address (0).
•You can use the default channel numbers (1 through 4) for both switch modules. Since each SCU is on a different port, and each port is addressed independently, there is no channel number conflict.
Switch Control Unit
Each switch control unit (SCU) provides bias and switch control of one switch module. Each SCU contains a decoder board, which allows you to set the SCU to a unique SCU address from 0 to 3 using DIP switches. The factory default setting is 0.
Switch Address
This is the complete software address for a specific switch. It is simply a concatenation of the SCU address (0, 1, 2, or 3) with the channel number (usually 1, 2, 3, or 4, but possibly a number up to 64).
Examples:
@2 = SCU address is 0 (and need not be specified), and channel 2 is selected.
@103 = SCU address 1 and channel 3 selected (usually a custom system).
@232 = SCU address is 2 and channel 32 is selected (custom systems only).
Switch Module
The
Typical
When a specification is noted as being typical it means that most units exhibit the stated performance, but not all. For this reason, the specification in not guaranteed by Agilent Technologies. Generally, typical specifications are provided for