Remote Programming Command Reference

Standard Command Reference

RUNTime:SWITch:DELay

 

Used for the delay between asserting the switch address list entry of the

 

RUNT:SWIT:SCAN command and the issuance of a receiver trigger. The

 

input to the command is in microseconds.

Default

The default is 2 microseconds.

Syntax

RUNT:SWIT:DEL <microseconds>;

Example

RUNT:SWIT:DEL 50; ! Set the trigger or setting

 

! time to 50 micro-seconds.

 

When using the multiple-parameter display mode of the 8530A, the

NOTE

 

minimum switch settling time setting is 50 s. Using faster settling times

 

can cause problems during the measurement.

 

For the Fast Data Acquisition modes, 2 s settling time is recommended.

 

 

7-12

85330A Multiple Channel Controller

Page 122
Image 122
Agilent Technologies 85330A service manual RUNTimeSWITchDELay