RUNTime:SWITch:SCAN

Set up the list of channel addresses for run-time control. For each event increment signal received by the 85330A, each channel address entry will be asserted and a subsequent trigger issued to begin a measurement of each data point.

This command can also be used to set the TTL level of the rear panel AUX 2 connector during run-time mode.

Scan list for port 1 and port 2 are empty.

AUX 2 connector default is TTL LOW.

RUNT:SWIT:SCAN 1 2,(@ <switch address list>);

or

RUNT:SWIT:SCAN 3,(@0 1);

This command controls port AUX 2.0 sets the AUX 2 BNC to a TTL low. A1 sets AUX2 to a TTL high.

AUX 1 cannot be controlled in RUNTime mode.

RUNT:SWIT:SCAN 1,(@101,102,101,102);

RUNT:SWIT:SCAN 2,(@101,101,102,102);

RUNT:SWIT:SCAN 3,(@0);

Sets AUX 2 output to TTL LOW during run-time.

Switch address list requirements are the same as the ROUT:CLOS command. A switch address scan list can remain empty and no data will be written to the port. If a port is not empty, then an error will be issued if the scan lists for both ports do not contain the same number of entries.