Performance Verification

Verifying Remote Ports 1 and 2

Remote Port 1 and remote Port 2 Test Procedure

1.Download the 85330A drivers version A.03.00 or above.

2.When select instrument is displayed, press:

{85330A}, {More}, {TEST}, {More}

3.Connect the modified connector to REMOTE 1 on the rear panel and press:

{REM_1_2}

4. The display should read:

“85330A_144:REM1 0101 PASS, REM2 1100 FAIL”

5.Move the modified connector to REMOTE 2 on the rear panel and press:

{REM_1_2}

6. The display should read:

“85330A_144: REM1 1111 FAIL, REM2 0011 PASS”

2-16

85330A Multiple Channel Controller

Page 46
Image 46
Agilent Technologies 85330A service manual Remote Port 1 and remote Port 2 Test Procedure