Remote Programming Command Reference

Standard Command Reference

RUNTime:EVENt:TRIGger

TTL Mode

IMM Mode

Default

Syntax

Example

The event trigger source allows either TTL triggering to control the run-time sequencing or IMMediate triggering to occur.

In TTL mode, the run-time control will:

1.Wait for an event trigger.

2.When the event trigger occurs, run-time control asserts each switch address entry in the scan list (one at a time), issuing a measurement trigger to the receiver for each.

3.Repeats step 2 for each frequency (defined using RUNT:SOUR:COUN).

Typically TTL mode is used for angle scans of CW or multi-frequency responses.

In IMM mode the 85330A issues the first trigger immediately on receiving the INIT:IMM command and continues on with the sequence. Typically, IMM mode is used for single angle frequency response measurements or timing measurements.

The default is TTL.

RUNT:EVENt:TRIGger TTL IMM;

OUTPUT 70918;”RUNT:EVEN:TRIG TTL;”

OUTPUT 70918;”RUNT:EVEN:TRIG IMM;”