Programming

To Use Run-Time Control Mode

Using More than One Controller

Starting Run-Time Mode

Run-Time

Measurement

Sequence

More than one multiple channel controller may be used in a system. Multiple controllers are usually used when the distance to a switch control unit exceeds the maximum length allowed. The REMOTE 1 and REMOTE 2 connectors on the rear panel allow connections to multiple controllers. See Chapter 9, “Service” for more information on these connectors.

To set up and control the remote controller, use these commands:

RUNTime:CONTroller

RUNTime:TIMEout:REMote

RUNTime:SWITch:TRIGger

The remote controllers can be triggered using the commands:

RUNTime:SOURce:SOURCE1:TRIGger

RUNTime:SOURce:SOURCE2:TRIGger

Issue the RUNT:INIT:IMM command to initiate the run-time operation. This command passes the hardware control to the 85330A and allows it to accept and issue triggers.

OUTPUT 70918;"RUNT:INIT:IMM;”

Once the sequence is initiated the 85330A either:

Waits for an event trigger before issuing a trigger to the receiver,

OR:

Immediately issues a trigger to the receiver.

This is set using the RUNT:EVEN:TRIG command.

Refer to Figure 6-2,“Run-Time Flow Chart with Programming Command References,” on page 6-16.The chart shows the sequence of a run-time measurement. Programming commands applicable to each function are shown. All programming commands must be executed before run-time is initiated. Here is a description of the run-time measurement sequence:

Idle

This is the idle state of the 85330A. This also the state the unit enters if an error occurs during run-time. Run-time mode is started when RUNT:INIT:IMM is executed.

Select Switch State

The first switch state is selected for Port 1 or Port 2, as defined with the RUNT:SWIT:SCAN command. A delay now occurs to allow the switch to settle, defined with RUNT:SWIT:DEL.