Programming

Programming Examples

Using Ground Mode

The following example uses the ground mode.

Figure 4-4

Ground Mode Example

Setup:

• Channel configuration mode: Auto

 

• Connection rule: Single. Multiple connection is available for the ground port.

 

• Connection sequence: Break_Before_Make

 

• Used module: All switch modules installed in the mainframe.

 

• Ground port: AUX Input 12

 

• Ground enabled input port: SMU Input 5 to 8

 

• Ground enabled output port: All output ports

 

• Connection paths (connected when the ground mode is ON):

 

from ground port to all ground enabled input/output ports

 

• Connection paths (connected by the :ROUT:CLOS command):

 

The following output ports will be disconnected from the ground port, and the

 

following input-output connections will be made.

 

from SMU1 to Output 1 (channel list 00101)

 

from SMU2 to Output 2 (channel list 00202)

 

from SMU3 to Output 3 (channel list 00303)

 

from SMU4 to Output 4 (channel list 00404)

4-12

Agilent B2200 User’s Guide, Edition 2

Page 110
Image 110