SCPI Command Reference

 

 

[:ROUTe]:AGND[:STATe]

 

 

 

[:ROUTe]:AGND[:STATe]

 

 

You cannot set the Ground Mode to ON when the Bias Mode is ON.

NOTE

 

 

 

This command controls the Ground Mode for the specified card. When the Ground

 

 

Mode is ON, the input Ground Port is connected to the all ground-enabledinput

 

 

ports/output ports that have not been connected to any other port. At *RST, the

 

 

Ground Mode is OFF.

 

 

 

The query returns the mode status.

 

 

The input Ground Port is set by “[:ROUTe]:AGND:PORT”. The ground-enabled

 

 

input ports are set by “[:ROUTe]:AGND:UNUSED”. And the ground-enabled

 

 

output ports are set by “[:ROUTe]:AGND:CHANnel:ENABle[:LIST]” or

 

 

“[:ROUTe]:AGND:CHANnel:ENABle:CARD”.

 

 

When the Ground Mode is ON, you cannot directly control the Ground Port

 

 

connections. However, they can be indirectly controlled when the connections of the

 

 

other input port are controlled.

 

 

When the Ground Mode is OFF, the input Ground Port is the same as the other input

 

 

ports, so relays can be controlled directly to connect to output ports.

Syntax

 

[:ROUTe]:AGND[:STATe] card_number,state

 

 

[:ROUTe]:AGND[:STATe]? card_number

 

 

 

 

 

 

 

Parameter

 

Explanation

 

 

 

 

 

 

 

card_number

 

For Auto Config mode: 0 or ALL

 

 

 

 

For Normal Config mode: 1, 2, 3, 4, or ALL

 

 

 

 

ALL is not available for query.

 

 

state

 

ON or 1: sets the Ground Mode ON

 

 

 

 

OFF or 0: sets the Ground Mode OFF

Query response

 

state <newline><^END>

 

 

 

0: OFF or 1: ON

 

Example

 

OUTPUT @Agb2200;":ROUT:AGND:STAT 0,ON"

 

 

OUTPUT @Agb2200;":ROUT:AGND:STAT? 0"

 

 

ENTER @Agb2200;A

 

In this example, A will be 1.