Programming

Programming Examples

Using Bias Mode

The following example uses the bias mode.

Figure 4-3

Bias Mode Example

Setup:

• Channel configuration mode: Auto

 

• Connection rule: Single. Multiple connection is available for the bias port.

 

• Connection sequence: Break_Before_Make

 

• Used module: All switch modules installed in the mainframe.

 

• Bias port: AUX Input 10

 

• Bias enabled output port: All output ports

 

• Connection paths (connected when the bias mode is ON):

 

from bias port to all bias enabled output ports

 

• Connection paths (connected by the :ROUT:CLOS command):

 

The following output ports will be disconnected from the bias port, and the

 

following input-output connections will be made.

 

from SMU1 to Output 1 (channel list 00101)

 

from SMU2 to Output 2 (channel list 00202)

 

from SMU3 to Output 3 (channel list 00303)

 

from SMU4 to Output 4 (channel list 00404)