This section describes how to connect the Agilent B2200 outputs to prober, connector plate, test fixture, and so on (DUT interface).

“Output Connectors”

“Connector Plates”

“To Make Connections to DUT Interface”

“To Make Interlock Circuit”

“To Mount Connectors”

Output Connections

If you do not use the connector plate for the connection between the output and the DUT interface, see “To Mount Connectors” on page 2-18.

Turn off the Agilent B2200 and all instruments connected. And do not turn them on until the connection described in this section is completed. If you ignore this warning, you may be exposed to dangerous voltage.

Output Connectors

The Agilent B2200 output connectors are the triaxial BNC connector. The input signals appears at the output as shown in Figure 2-1 when input-output switching path is made. Figure 2-1 shows signals appear on the output connectors with non-Kelvin connection. If you make Kelvin connection, the signal of the even output connectors is Sense, not Force.

Output Connector and Output Signal

Signal (force)

Signal (force)

GUARD

GND

GND

(a) When SMU is connected

(b) When AUX is connected

Agilent B2200 User’s Guide, Edition 2